Parts · Cooling filtration
DuPure DP-808-IPL
2-in-1 dual-stage cooling filter for IPL and 808 nm diode laser equipment (aesthetic laser systems). One compact housing combines a high-efficiency PP cotton particulate pre-filter with deionizing / softening resin, replacing conventional bulky dual-canister setups with a single unit.
Price on request
- Model
- DP-808-IPL
- Product
- DuPure 2-in-1 Dual-Stage Filter
- Application
- Aesthetic laser systems — IPL & 808 nm diode laser equipment
- Housing material
- Medical-grade polypropylene (PP) (wording from the product manual; not an independent certification claim)
- Internal filtration media
- Melt-blown PP & cation resin
- Filtration rating
- 5 micron (particle retention)
- Max. working pressure
- 120 PSI (8.27 bar)
- Flow capacity
- 5 L/min (continuous circulation)
- Operating temperature
- 20 °C – 30 °C (68 °F – 86 °F)
- Service lifetime
- 6 months (or per DI alarm)
- Applicable coolant
- Distilled, deionized (DI), purified
Core advantages
Dual-stage protection
Outer PP cotton stops particles down to 5 microns, protecting pump impellers and micro-channels. The internal ion-exchange resin extracts Ca²⁺ and Mg²⁺, preventing limescale encrustation on laser emitters.
Space-saving footprint
The integrated concentric design eliminates secondary filter brackets, inter-stage tubing and extra fittings, freeing internal cabinet space and reducing potential leak points.
Cost & maintenance efficient
Particulate and softening media are replaced together in one cartridge, cutting routine service time and optimizing circuit hydraulic resistance.
Installation & operating guidelines
- Flow direction: Verify the inline direction before connection. Coolant must enter via the marked “IN” port and discharge through the “OUT” port.
- Approved coolant: Always charge the cooling system with high-grade deionized (DI) water or distilled water. Never use unfiltered tap water or mineral-rich liquids.
- Air purging & leak testing: After seating push-fit or threaded connections, cycle the coolant pump at low speed to bleed air bubbles before powering the diode driver.
- Scheduled replacement: Replace the cartridge every 6 months, or immediately if the machine raises a water conductivity or flow-rate warning.
FAQ
What does the DuPure DP-808-IPL filter do?
It integrates a 5-micron PP cotton particulate pre-filter and deionizing / softening (cation) resin in one housing, for the cooling water of IPL and 808 nm diode laser equipment. It replaces bulkier dual-canister filter setups with a single unit.
Which coolant should I use?
Distilled, deionized (DI) or purified water. Never use unfiltered tap water or mineral-rich liquids.
When should the cartridge be replaced?
Every 6 months, or immediately if the machine raises a water conductivity or flow-rate warning (DI alarm).
What are the operating limits?
Maximum working pressure 120 PSI (8.27 bar); continuous flow capacity 5 L/min; operating temperature 20–30 °C (68–86 °F).
How do I install it correctly?
Connect coolant to the marked IN port and take it out from the OUT port. After seating the connections, run the coolant pump at low speed to bleed air before powering the diode driver.
Is “Medical-grade” a certification?
The product manual lists the housing material as “Medical-grade Polypropylene (PP)”. This page repeats that wording as a material description and does not claim any independent certification or regulatory clearance.